Σάββατο 14 Απριλίου 2018

Experimental validation of robot-assisted cardiovascular catheterization: model-based versus model-free control

Abstract

Purpose

In cardiac electrophysiology, a long and flexible catheter is delivered to a cardiac chamber for the treatment of arrhythmias. Although several robot-assisted platforms have been commercialized, the disorientation in tele-operation is still not well solved. We propose a validation platform for robot-assisted cardiac EP catheterization, integrating a customized MR Safe robot, a standard clinically used EP catheter, and a human–robot interface. Both model-based and model-free control methods are implemented in the platform for quantitative evaluation and comparison.

Methods

The model-based and model-free control methods were validated by subject test (ten participants), in which the subjects have to perform a simulated radiofrequency ablation task using both methods. A virtual endoscopic view of the catheter is also provided to enhance hand-to-eye coordination. Assessment indices for targeting accuracy and efficiency were acquired for the evaluation.

Results

(1) Accuracy: The average distance measured from catheter tip to the closest lesion target during ablation of model-free method was 19.1% shorter than that of model-based control. (2) Efficiency: The model-free control reduced the total missed targets by 35.8% and the maximum continuously missed targets by 46.2%, both indices corresponded to a low p value ( \(\le 0.05\) ).

Conclusion

The model-free method performed better in terms of both accuracy and efficiency, indicating the model-free control could adapt to soft interaction with environment, as compared with the model-based control that does not consider contacts.



from Imaging via alkiviadis.1961 on Inoreader https://ift.tt/2qwtTfg

Δεν υπάρχουν σχόλια:

Δημοσίευση σχολίου

Δημοφιλείς αναρτήσεις