Simulation study of a D-shape PET scanner for improved sensitivity and reduced cost in whole-body imaging

Much research effort is being made to increase the sensitivity and improve the imaging performance of positron emission tomography (PET) scanners. Conventionally, sensitivity can be increased by increasing the number of detector rings in the axial direction (but at high cost) or reducing the diameter of the scanner (with the disadvantages of reducing the space for patients and degrading the spatial resolution due to the parallax error). In this study, we proposed a PET scanner with a truncated ring and an array of detectors that can be arranged in a straight line below the bed. We called this system ‘D-PET’ as it resembles the letter ‘D’ when it is rotated by 90° in the counterclockwise direction. The basic design idea was to cut the unused space under the patient’s bed; this area is usually not in use in clinical diagnosis. We conducted Monte Carlo simulations of the D-PET scanner and compared its performance with a cylindrical PET scanner. The scanners were constructed from 4-...