Impact of Contrast Media Concentration on Low-Kilovolt Computed Tomography Angiography: A Systematic Preclinical Approach

imageObjectives Low peak kilovoltage (kVp) protocols in computed tomography angiography (CTA) demand a review of contrast media (CM) administration practices. The aim of this study was to systematically evaluate different iodine concentrations of CM in a porcine model. Materials and Methods Dynamic 70 kVp CTA was performed on 7 pigs using a third-generation dual-source CT system. Three CM injection protocols (A-C) with an identical total iodine dose and iodine delivery rate (150 mg I/kg, 12 s, 0.75 g I/s) differed in iodine concentration and flow rate (protocol A: 400 mg I/mL, 1.9 mL/s; B: 300 mg I/mL, 2.5 mL/s; C: 150 mg I/mL, 5 mL/s). All protocols were applied in a randomized order and compared intraindividually. Arterial enhancement at different locations in the pulmonary artery, the aorta, and aortic branches was measured over time. Time attenuation curves, peak enhancement, time to peak, and bolus tracking delay times needed for static CTA were calculated. The reproducibility of optimal parameters was tested in single-phase CTA. Results The heart rates of the pigs were comparable for all protocols (P > 0.7). The injection pressure was significantly higher for protocol A (64 ± 5 psi) and protocol C (55 ± 3 psi) compared with protocol B (39 ± 2 psi) (P