Changes in the ADC of diffusion-weighted MRI with the oscillating gradient spin-echo (OGSE) sequence due to differences in substrate viscosities

Abstract

Purpose

Compared with the conventional pulsed gradient spin-echo (PGSE) sequence, diffusion-weighted imaging (DWI) with the oscillating gradient spin-echo (OGSE) sequence can shorten the diffusion time by changing the frequency. The purpose was to investigate whether n-alkanes are suitable as isotropic phantoms for estimating the diffusion coefficient with the OGSE sequence.

Materials and methods

We investigated changes in the apparent diffusion coefficient (ADC) due to differences in the viscosities of nine n-alkane phantoms with different numbers of carbon atoms from C8H18 to C16H34 using OGSE and PGSE sequences at 21 °C. Effective diffusion times of 4.3, 5.1, 6.5, 9.3, 20, 40, and 60 ms were used. The T2 relaxation times of each n-alkane phantom were measured using quantitative synthetic magnetic resonance imaging (MRI). Circular regions of interest were placed manually within the alkane phantoms on ADC and T2 maps.

Results

In each alkane phantom, changes in mean ADC values were almost constant with changes in diffusion times. Viscosities and ADC values showed inverse proportionality, as expected theoretically.

Conclusion

The ADC values of alkanes do not depend on diffusion times. The n-alkanes can be useful phantoms for assessing the accuracy of clinical protocols of DWI with the OGSE sequence.



from Imaging via alkiviadis.1961 on Inoreader https://ift.tt/2vMtaf7