by Jonathan M. Bernstein, Michael J. Daly, Harley Chan, Jimmy Qiu, David Goldstein, Nidal Muhanna, John R. de Almeida, Jonathan C. Irish
BackgroundWe set out to determine the accuracy of 3D-navigated mandibular and maxillary osteotomies with the ultimate aim to integrate virtual cutting guides and 3D-navigation into ablative and reconstructive head and neck surgery.
MethodsFour surgeons (two attending, two clinical fellows) completed 224 unnavigated and 224 3D-navigated osteotomies on anatomical models according to preoperative 3D plans. The osteotomized bones were scanned and analyzed.
ResultsMedian distance from the virtual plan was 2.1 mm unnavigated (IQR 2.6 mm, ≥3 mm in 33%) and 1.2 mm 3D-navigated (IQR 1.1 mm, ≥3 mm in 6%) (P Conclusion
3D-rendering enables osteotomy navigation. 3 mm is an appropriate planning distance. The next steps are translating virtual cutting guides to free bone flap reconstruction and clinical use.