A procedure to estimate the activity in target tissues in patients during the therapeutic administration of 131 I radiopharmaceutical treatment for thyroid conditions (hyperthyroidism and differentiated thyroid cancer) using a gamma camera (GC) with a high energy (HE) collimator, is proposed. Planar images are acquired for lesions of different sizes r , and at different distances d , in two HE GC systems. Defining a region of interest (ROI) on the image of size r , total counts n g are measured. Sensitivity S (cps MBq −1 ) in each acquisition is estimated as the product of the geometric G and the intrinsic efficiency η 0 . The mean fluence of 364 keV photons arriving at the ROI per disintegration G , is calculated with the MCNPX code, simulating the entire GC and the HE collimator. Intrinsic efficiency η 0 is estimated from a calibration measurement of a plane reference ...