Publication date: Available online 26 March 2018
Source:Academic Radiology
Author(s): Takanori Masuda, Yoshinori Funama, Takeshi Nakaura, Tomoyasu Satou, Tomokazu Okimoto, Yukari Yamashita, Naoyuki Imada, Kazuo Awai
Rationale and ObjectivesWe compared the radiation dose and diagnostic accuracy on 120- and 100-kVp coronary computed tomography angiography (CCTA) scans whose contrast-to-noise ratio (CNR) was the same.Materials and MethodsWe studied 1311 coronary artery segments from 100 patients. For 120-kVp scans, the targeted image level was set at 25 Hounsfield units (HU). For 100-kVp scans, the targeted noise level was set at 30 HU to obtain the same CNR as at 120 kVp. We compared the CNR and the radiation dose on scans acquired at 120 and 100 kVp. Invasive coronary angiography (ICA) images were evaluated by an interventional coronary angiography specialist, and CCTA images were evaluated by a radiologist. Coronary artery disease was defined as a luminal narrowing ≧50% for ICA and CCTA. With ICA considered the gold standard, the diagnostic accuracy (sensitivity, specificity, positive predictive value, and negative predictive value) was analyzed on both 120- and 100-kVp CCTA images. We also compared the diagnostic accuracy for area under the receiver operating characteristic curve of the ICA and CCTA performed at 120 and 100 kVp. Two blinded observers visually evaluated the septal branch.ResultsThe mean dose-length product was 48% lower at 100 kVp than at 120 kVp (P < .01). Under the 120-kVp CCTA protocol, the area under the curve, 95% confidence interval, sensitivity, specificity, positive predictive value, and negative predictive value were 0.94%, 0.91%–0.96%, 94.0%, 93.0%, 82.3%, and 98.1%, respectively; at 100 kVp these values were 0.94%, 0.92%–0.97%, 96.1%, 92.0%, 85.2%, and 98.0%, respectively. Area under the receiver operating characteristic curve analysis revealed no significant difference in diagnostic accuracy between the two protocols (P = .87).ConclusionsAt the same CNR, the 100-kVp CCTA protocol may help to reduce the radiation dose by approximately 50% compared to the 120-kVp protocol without degradation of diagnostic accuracy.
from Imaging via alkiviadis.1961 on Inoreader https://ift.tt/2Gva0i1
Εγγραφή σε:
Σχόλια ανάρτησης (Atom)
Δημοφιλείς αναρτήσεις
-
Abstract Objective To evaluate Chinese medicine (CM) formula Bazheng Powder (八正散) as an alternative therapeutic option for female patients...
-
To evaluate the effect of Recurrence Score® results (RS; Oncotype DX® multigene assay ODX) on treatment recommendations by Swiss multidiscip...
-
Abstract Soil conditioners can be used to compensate for the insufficient soil nutrition and organic matter (OM) of arable soils. However, ...
-
Ocular Vestibular Evoked Myogenic Potentials: Where Are We Now? Objective: Over the last decade, ocular vestibular evoked myogenic potential...
-
Abstract Objective To study the effects of Astragalus polysaccharide (APS), the primary effective component of the Chinese herb medicine A...
-
Geriatric trauma: A population-based study Saint Shiou-Sheng Chen, Li-Chien Chien Formosan Journal of Surgery 2019 52(2):39-44 Background: G...
-
Objectives Adult sagittal posture is established during childhood and adolescence. A flattened or hypercurved spine is associated with poore...
-
Related Articles Screening for Atrial Fibrillation using Economical and accurate TechnologY (SAFETY)-a pilot study. BMJ Open. 2017 Ja...
-
Abstract Purpose Overcoming the flaws of current data management conditions in head and neck oncology could enable integrated informatio...
Δεν υπάρχουν σχόλια:
Δημοσίευση σχολίου