Prototype volumetric ultrasound tomography image guidance system for prone stereotactic partial breast irradiation: proof-of-concept

Accurate dose delivery in stereotactic partial breast irradiation (S-PBI) is challenging because of the target position uncertainty caused by breast deformation, the target volume changes caused by lumpectomy cavity shrinkage, and the target delineation uncertainty on simulation computed tomography (CT) images caused by poor soft tissue contrast. We have developed a volumetric ultrasound tomography (UST) image guidance system for prone position S-PBI. The system is composed of a novel 3D printed rotation water tank, a patient-specific resin breast immobilization cup, and a 1D array ultrasound transducer. Coronal 2D US images were acquired in 5° increments over a 360° range, and planes were acquired every 2 mm in elevation. A super-compounding technique was used to reconstruct the image volume. The image quality of UST was evaluated with a BB-1 breast phantom and BioZorb surgical marker, and the results revealed that UST offered better soft tissue contrast than CT and similar ima...