Verification of I-125 brachytherapy source strength for use in radioactive seed localization procedures.

http:--linkinghub.elsevier.com-ihub-imag Related Articles

Verification of I-125 brachytherapy source strength for use in radioactive seed localization procedures.

Appl Radiat Isot. 2016 06;112:62-8

Authors: Metyko J, Erwin W, Landsberger S

Abstract
A general-purpose nuclear medicine dose calibrator was assessed as a potential replacement for a dedicated air-communicating well-type ionization chamber (brachytherapy source strength verification instrument) for (125)I seed source strength verification for radioactive seed localization, where less stringent accuracy tolerances may be acceptable. The accuracy, precision and reproducibility of the dose calibrator were measured and compared to regulatory requirements. The results of this work indicate that a dose calibrator can be used for (125)I seed source strength verification for radioactive seed localization.

PMID: 27015651 [PubMed - indexed for MEDLINE]