Τρίτη 13 Φεβρουαρίου 2018

Calibration setting numbers for dose calibrators for the PET isotopes (52)Mn, (64)Cu, (76)Br, (86)Y, (89)Zr, (124)I.

http:--linkinghub.elsevier.com-ihub-imag Related Articles

Calibration setting numbers for dose calibrators for the PET isotopes (52)Mn, (64)Cu, (76)Br, (86)Y, (89)Zr, (124)I.

Appl Radiat Isot. 2016 07;113:89-95

Authors: Wooten AL, Lewis BC, Szatkowski DJ, Sultan DH, Abdin KI, Voller TF, Liu Y, Lapi SE

Abstract
For PET radionuclides, the radioactivity of a sample can be conveniently measured by a dose calibrator. These devices depend on a "calibration setting number", but many recommended settings from manuals were interpolated based on standard sources of other radionuclide(s). We conducted HPGe gamma-ray spectroscopy, resulting in a reference for determining settings in two types of vessels containing one of several PET radionuclides. Our results reiterate the notion that in-house, experimental calibrations are recommended for different radionuclides and vessels.

PMID: 27152914 [PubMed - indexed for MEDLINE]



from Imaging via alkiviadis.1961 on Inoreader http://ift.tt/2o2PZ7i

Δεν υπάρχουν σχόλια:

Δημοσίευση σχολίου

Δημοφιλείς αναρτήσεις