Δευτέρα 5 Φεβρουαρίου 2018

High-pitch low-voltage CT coronary artery calcium scoring with tin filtration: accuracy and radiation dose reduction

Abstract

Objectives

To investigate diagnostic accuracy and radiation dose of high-pitch CT coronary artery calcium scoring (CACS) with tin filtration (Sn100kVp) versus standard 120kVp high-pitch acquisition.

Methods

78 patients (58% male, 61.5±9.1 years) were prospectively enrolled. Subjects underwent clinical 120kVp high-pitch CACS using third-generation dual-source CT followed by additional high-pitch Sn100kVp acquisition. Agatston scores, calcium volume scores, Agatston score categories, percentile-based risk categorization and radiation metrics were compared.

Results

61/78 patients showed coronary calcifications. Median Agatston scores were 34.9 [0.7–197.1] and 41.7 [0.7–207.2] and calcium volume scores were 34.1 [0.7–218.0] for Sn100kVp and 35.7 [1.1–221.0] for 120kVp acquisitions, respectively (both p<0.0001). Bland-Altman analysis revealed underestimated Agatston scores and calcium volume scores with Sn100kVp versus 120kVp acquisitions (mean difference: 16.4 and 11.5). However, Agatston score categories and percentile-based risk categories showed excellent agreement (ĸ=0.98 and ĸ=0.99). Image noise was 25.8±4.4HU and 16.6±2.9HU in Sn100kVp and 120kVp scans, respectively (p<0.0001). Dose-length-product was 9.9±4.8mGy*cm and 40.9±14.4mGy*cm with Sn100kVp and 120kVp scans, respectively (p<0.0001). This resulted in significant effective radiation dose reduction (0.13±0.07mSv vs. 0.57±0.2mSv, p<0.0001) for Sn100kVp acquisitions.

Conclusion

CACS using high-pitch low-voltage tin-filtered acquisitions demonstrates excellent agreement in Agatston score and percentile-based cardiac risk categorization with standard 120kVp high-pitch acquisitions. Furthermore, radiation dose was significantly reduced by 78% while maintaining accurate risk prediction.

Key points

• Coronary artery calcium scoring with tin filtration reduces radiation dose by 78%.

• There is excellent correlation between high-pitch Sn100kVp and standard 120kVp acquisitions.

• Excellent agreement regarding Agatston score categories and percentile-based risk categorization was achieved.

• No cardiac risk reclassifications were observed using Sn100kVp coronary artery calcium scoring.



from Imaging via alkiviadis.1961 on Inoreader http://ift.tt/2BVbnQT