Evaluation of PET Scanner Performance in PET/MR and PET/CT Systems: NEMA Tests.
Mol Imaging Radionucl Ther. 2018 Feb 01;27(1):10-18
Authors: Demir M, Toklu T, Abuqbeitah M, Çetin H, Sezgin HS, Yeyin N, Sönmezoğlu K
Abstract
OBJECTIVE: The aim of the present study was to compare the performance of positron emission tomography (PET) component of PET/computed tomography (CT) with new emerging PET/magnetic resonance (MR) of the same vendor.
METHODS: According to National Electrical Manufacturers Association NU2-07, five separate experimental tests were performed to evaluate the performance of PET scanner of General Electric GE company; SIGNATM model PET/MR and GE Discovery 710 model PET/CT. The main investigated aspects were spatial resolution, sensitivity, scatter fraction, count rate performance, image quality, count loss and random events correction accuracy.
RESULTS: The findings of this study demonstrated superior sensitivity (~ 4 folds) of PET scanner in PET/MR compared to PET/CT system. Image quality test exhibited higher contrast in PET/MR (~ 9%) compared with PET/CT. The scatter fraction of PET/MR was 43.4% at noise equivalent count rate (NECR) peak of 218 kcps and the corresponding activity concentration was 17.7 kBq/cc. Whereas the scatter fraction of PET/CT was found as 39.2% at NECR peak of 72 kcps and activity concentration of 24.3 kBq/cc. The percentage error of the random event correction accuracy was 3.4% and 3.1% in PET/MR and PET/CT, respectively.
CONCLUSION: It was concluded that PET/MR system is about 4 times more sensitive than PET/CT, and the contrast of hot lesions in PET/MR was ~ 9% higher than PET/CT. These outcomes also emphasize the possibility to achieve excellent clinical PET images with low administered dose and/or a short acquisition time in PET/MR.
PMID: 29393048 [PubMed]
from Imaging via alkiviadis.1961 on Inoreader http://ift.tt/2s5vNae
Δεν υπάρχουν σχόλια:
Δημοσίευση σχολίου